The CCI PiMPro 1921Rm is a Rack Mount precision Passive Intermod (PIM) analyzer that has been designed to meet the specific challenges of PIM testing both in laboratory and production environments. It covers both the 1850-1990 MHz and 1710-2155 MHz bands. The analyzer delivers maximum power of two carriers at 40 watts (40W x 2) and accuracy of intermodulation (IM) measurements, with a sensitivity of -125 dBm( -168 dBc at 20 watts). In a convenient 19 inch rack mount configuration. The PiMPro provides precise measurement of the 3rd, 5th, 7th and 9th order of intermodulation of any system or component under high-power conditions. In addition to passive intermodulation measurements, the unit will provide VSWR and Return Loss values.
Band: PCS 1900 / AWS 2100. Receive Frequency: 1850 - 1910 MHz / 1710 - 1755 MHz. Transmit Frequency: 1930 - 1990 MHz / 2110 - 2155 MHz. RF Output Connector: 7-16 DIN Female.
Transmitter; Frequency Accuracy: < 5 ppm. Power Accuracy: 0.3 dB. Frequency Step Size: 200 kHz. Power Resolution: 0.1 dB. Adjustable Power Range: 17 to 46 dBm × 2 (50 mW to 40 W × 2).
Receiver; Residual Intermod Level: -122 dBm (-125 dBm Typical). Measurement Sensitivity: -135 dBm. Noise Floor: -136 dBm. Reverse Power Protection: 13 dBm (20 mW) continuous.
Measurement Mode; Measurement Method: One Port, Reverse PIM. Real Time PIM: 3rd, 5th, 7th & 9th PIM. Return Loss: Measured in dB. PIM vs Time: 3rd, 5th, 7th & 9th PIM. RX Interference: Receive Only Mode - Noise Floor Measurements. Frequency Sweep: Frequency Response. DAS Feature: Path Loss Characteristics.
|